Sirona Dental Systems GmbH
11 Transversal slices (TSA)
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
11.1 Preparing a TSA exposure
59 87 594 D 3352
D 3352.201.01.18.02
87
11
Transversal slices (TSA)
11.1
Preparing a TSA exposure
In combination with a panoramic slice view, transversal
slices enable a 3-dimensional representation of the dif-
ferent maxillary and mandibular regions.
•
Switch on the unit and make SIDEXIS ready for ex-
posure.
Plug in TSA sensor
•
Make sure that a TSA sensor is plugged into the PAN
slot –
either
PAN + TSA
or
CEPH + TSA
ATTENTION
DO NOT DROP THE SENSOR!
A shock sensor for detecting shocks or drops is built in.
TSA application
TSA imaging is used for example for implant insertion
(A) or displaced teeth (B).
Ball positioning
Prior to the exposure you must fix a steel ball (e.g. with
a diameter of 5 mm) at the planned implant placement
point by means of elastic impression wax.
(This step is not required for a displaced tooth)
T
R
CEPH
N
PAN
TS
TS
LS
LS
Filmkassette einrasten
62kV
8mA
?
P6.1
12,1s
CEPH
CEPH
PAN
AN
TS
P6
12,8s
P
AN +
TSA
CEPH +
TSA
B
A