Sirona Dental Systems GmbH
9 Operation
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
9.9 Remote control
59 87 594 D 3352
D 3352.201.01.18.02
71
9.9
Remote control
If the ORTHOPHOS XG is located in an X-ray room
which has a door and enables visual contact with the
patient, you can use
remote control
to release the
exposure.
For that purpose, the
exposure release button (10)
can
be detached from the unit and attached to the remote
control.
The
exposure release button (A)
can be used if a lon-
ger cable is not required to maintain visual contact with
the patient.
The remote control has an
“R” key
(36) for acknowledg-
ing the exposure and resetting the unit to its starting
position, an optical (32) and acoustic
radiation indica-
tor
as well as a “
Unit ON
” LED (33).
After you switch on the system, the LED (33) lights up.
The radiation indicator (32) lights up for approx. one sec-
ond for a function test.
The four display fields on the display panel (B) light up
and program P1 appears with the corresponding values
after a short time.
As long as plain-text
help messages
are displayed on
the Easypad touchscreen, they also appear in coded
form on the
"Prog."
display field of the remote control,
alternating
with the
program name
.
Once the unit has been prepared for the exposure and
all
help messages
have been processed, the
program
name “Prog.”
, the
exposure time “s”
as well as the
“
kV
” and “
mA
” values are constantly displayed on the
display panel (B).
You may now release the exposure.
NOTE
i
If a row of dots
........
appears in the
Prog.
field, this
means that the system is currently in a preparatory
phase (e.g. system movements, parameter changes,
program loading times etc.). Just wait until the dots auto-
matically disappear and the system signals that it is
ready again.
R
s
k
V
m
A
R
TM 3
8.1
71
8
TM 3
8.1
71
8
Prog.
s
k
V
m
A
Prog.
..........
8.1
s
Prog.
A
10
36
32
33
B