Taking out of operation and disposal
12.2 Scrapping the analyzer
ULTRAMAT 23
182
Manual, 3/2016, A5E37100388-003
12.2
Scrapping the analyzer
If the ULTRAMAT 23 is to be scrapped, take it of operation as follows:
Rack unit
1.
Make sure that gas is no longer flowing through the analyzer. If external pumps are
present, switch all of them off.
2.
Purge the sample gas path with air or nitrogen.
3.
Switch the analyzer off.
4.
Disconnect the power plug.
5.
Disconnect all hose connections from the rear of the analyzer.
With pipe versions, unscrew all pipes.
6.
In the case of analyzers with an electrochemical oxygen sensor, remove this from the
analyzer (see section Replacing the electrochemical oxygen sensor (Page 171)).
7.
In the case of analyzers with a hydrogen sulfide sensor, remove this from the analyzer
(see section Replacing the hydrogen sulfide sensor (Page 172)).
Bench-top unit
1.
Make sure that gas is no longer flowing through the analyzer. If external pumps are
present, switch all of them off.
2.
Purge the sample gas path with air or nitrogen.
3.
Switch the analyzer off.
4.
Disconnect the power plug.
5.
Empty the condensation trap (see section Emptying the condensation trap (Page 170)).
6.
Disconnect the supply hose from the condensation trap.
7.
Disconnect all hose connections from the rear of the analyzer.
Product disposal
The analyzer to be disposed of as electronic waste with the code number 160213 is a
'product containing dangerous components'. It must therefore be disposed of correctly and in
an environmentally-friendly manner by a local waste disposal company.
Disposal of the electrochemical oxygen sensor
The exhausted or faulty O
2
sensor is hazardous waste and must be packed and disposed of
accordingly!
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...