Functions
8.1 Diagnostics
ULTRAMAT 23
110
Manual, 3/2016, A5E37100388-003
•
Reference/bridge
–
Reference: The reference voltage for the electronics of
the analyzer.
–
Bridge: The supply voltage to the measuring bridge.
•
LCD temp.
The temperature which determines the display contrast.
Setting of the LCD contrast is described in section
Parameter: Pump/LCD contrast (Page 130).
•
Analog output
The actual value of the output current is displayed in mA for
each of the measured components (the dimension is not
shown for space reasons) as well as the start-of-scale value
(either 0, 2 or 4 mA) and the full-scale value (20 mA) of the
output current range. Refer to section Configuration:
Inputs/outputs/pump: Analog outputs (Page 134) for
adjustment of the start-of-scale values.
8.1.3
Analyzer status: Factory settings hardware
Factory settings are parameters which are already set on de-
livery such as
•
Serial No.
•
Revision
•
Order No.
The hardware configuration and release version can be read
here.
8.1.4
Analyzer status: Factory settings software
Factory settings are parameters which are already set on de-
livery such as
•
Software/firmware release version
The software release version can be read here.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...