Service and maintenance
10.2 Maintenance work
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
169
To replace the fuses, proceed as follows:
1.
Remove the fuse holder above the appliance plug.
Use a small screwdriver to do this.
2.
Remove the blown fuse from the holder.
3.
Insert a new fuse.
4.
Insert the fuse holder into the compartment again.
Note
Only fuses of the type printed on the rear of the analyzer may be used (see also
section Electronics (Page 190)).
10.2.5
Replacing the fine safety filter
To replace the filter, proceed as follows:
●
Unscrew the four screws of the top housing cover, and remove the cover to the rear.
●
Determine the contaminated filter according to its type (see Gas path (Page 187)).
●
Remove the hoses from the filter.
●
Remove the old filter.
The filter must be disposed of as residual waste.
●
Insert the new filter.
When installing the filter, make sure that the arrow on the filter points in the gas flow
direction.
●
Push the cover back onto the housing and screw tight.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...