Description
3.4 Technical specifications
ULTRAMAT 23
36
Manual, 3/2016, A5E37100388-003
Influencing variables
Drift
•
With AUTOCAL
•
Without AUTOCAL
Negligible
< 2 % of smallest measuring range/week
Temperature
max. 2 % of smallest possible measuring range according to label per
10 K with an AUTOCAL cycle time of 6 h
Air pressure
< 0.2 % of measuring range per 1 % change in pressure
Power supply
< 0.1 % of output signal span with a variation of ±10 %
Line frequency
± 2 % of full-scale value with a frequency variation
of ± 5 %
Deviations with measuring range 0 ... 200 mg/m
3
SO
2
(analyzer versions 7MB2335-xNBxx-xAAx, 7MB2337-xNBxx, 7MB2337-xxxxx-xNBx, 7MB2338-xxxxx-xNBx)
Availability
Max. 95 %
AUTOCAL cycle time
Max. 6 h
Temperature variations
Max. 1 °C (1.8 °F)
The device must not be operated in an area subject to drafts. This is
especially valid for the rear panel with large cooling element.
Other
This measuring range has not been suitability-tested.
Deviations with measuring range 0 ... 100 mg/m
3
SO
2
(analyzer versions 7MB2335-xNTxx-xAAx, 7MB2337-xNTxx-xxxx, 7MB2337-xxxxx-xNTx, 7MB2338-xxxxx-xNTx)
Delayed display (T
90
time)
Depending on the external gas conditioning.
Note: SO
2
is highly soluble in water
Output signal resolution
≤ 0.1
mg/m
3
Linearity error
≤ 2 mg/m³ (2% of smallest measuring range)
Drift
•
With constant ambient conditions
•
After temperature jump
≤ 0.33 mg/m³ per hour after a running
-in period of 200 hours
Not defined
Temperature influence
≤ 8
mg/m³ per 10 K dev
iation; additional drift after temperature jump
Influence of auxiliary power
≤
0.4 mg/m³ with a change of ±
10%
Influence of line frequency
≤
8 mg/m³ with a change of ±
5%
Cross-sensitivities
Refer to following table
Availability
Max. 95 %
AUTOCAL cycle time
Max. 3 h
Temperature variations
Max. 1 °C (1.8 °F)
The device must be operated at a constant ambient temperature. The
device must not be operated in an area subject to drafts. This is especial-
ly valid for the rear panel with large cooling element.
Other
This measuring range has not been suitability-tested.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...