Description
3.3 Function
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
27
generates a pneumatic connection between the upper layer and the lower layers. Negative
feedback from the upper and lower layers leads to an overall narrowing of the spectral
sensitivity band. The volume of the third layer and, therefore, the absorption of the bands,
can be varied using a slide (10), thereby increasing the selectivity of each individual
measurement.
The rotating chopper (5) generates a pulsating flow in the receiver chamber that the
microflow sensor (3) converts into an electrical signal. The microflow sensor consists of two
nickel-plated grids heated to approximately 120 ºC (248 °F), which, along with two
supplementary resistors, form a Wheatstone bridge. The pulsating flow together with the
dense arrangement of the nickel grids causes a change in resistance. This leads to an offset
in the bridge which is proportional to the concentration of sample gas.
Note
Contamination of the analyzer chambers
The sample gases must be fed into the analyzers free of dust. Condensation should also be
prevented in the analyzer chambers Therefore, the use of gas modified for the measuring
task is necessary in most application cases.
Furthermore, the ambient air of the sensor must not have large concentrations of the
components to be measured.
Electrochemical oxygen measurement
1
Gold cathode
4
Signal output
2
Electrolyte (acetic acid)
5
Lead anode
3
Thermistor and load resistor for tempera-
ture compensation
6
Oxygen diffusion membrane made of FEP
Figure 3-5
Operating principle of the electrochemical O
2
sensor
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...