Commissioning
6.2 Safety instructions
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
79
CSA Class I Div. 2 and ATEX Zone 2
The following safety and warning information applies to analyzers (special versions) which
are operated in accordance with CSA Class I Div. 2 (hazardous locations) and ATEX Zone 2:
WARNING
Potentially explosive atmosphere
Do not open, service or repair in an area in which a potentially explosive atmosphere may
be present.
ATEX Zone 2
The following applies to devices (special versions) which are operated according to ATEX in
Ex zone 2:
DANGER
Explosion hazard
The ULTRAMAT 23 gas analyzers for use in Ex zone 2 must be installed in an appropriate
enclosure. This enclosure must comply with the requirements of EN 60079-15 and must be
designed for all ambient conditions which can occur during operation.
The maximum ambient temperature is 50 °C.
Suitable measures must additionally be applied to ensure that
•
the generation of potentially explosive gas mixtures inside the analyzer does not exceed
the level of Zone 2
•
interferences cannot lead to a deviation of more than 40% from the rated voltage.
Note
In the case of device versions for use in Ex zone 2, it is also essential to observe the 'ATEX
compact operating instructions for rack units of Series 6' (A5E37888334)!
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...