Description
3.4 Technical specifications
ULTRAMAT 23
38
Manual, 3/2016, A5E37100388-003
3.4.4
Electrochemical oxygen sensor
General
Measuring ranges
0 ... 5% to 0 ... 25% O
2
, parameterizable
Associated gases
The oxygen sensor must not be used if the associated gas
contains the following components:
•
Chlorine or fluorine compounds
•
Heavy metals
•
Aerosols
•
Mercaptans
•
Alkaline components (e.g. NH
3
in % range)
Service life
Approx. 2 years with 21% O
2
Time response
Response time (T
90
time)
Dependent on dead time and parameterizable damping,
<30 s with sample gas flow of approx. 1.2 l/min
Measuring response
Output signal noise
< 0.5% of the full-scale value
Display resolution
< 0.2% of the full-scale value
Output signal resolution
< 0.2% of the output signal span
Reproducibility
≤ 0.05% O
2
Climatic conditions
Permissible ambient temperature
•
During operation
•
During transportation and storage
+5 ... +45 °C (41 ... 113 °F)
-20 ... +60 °C (-4 ... 140 °F)
Permissible ambient humidity
< 90% RH (relative humidity) during transportation and storage
Permissible ambient pressure
600 ... 1200 hPa absolute
Maximum operating altitude
2000 m above sea level
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...