ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
161
Application note
9
9.1
H2S sensor with 'small' measuring range
This section reflects the experiences gained with operation of a sensor for a measuring
range of 5/50 vpm H
2
S.
Packaging
The transport packaging is not gas-tight. To prevent drying-out of the sensor and the
possibility of malfunctions, the sensor may only be stored for a period up to 3 months.
Storage and transport
Storage at an excessively high humidity (tropics, several months) results in swelling of the
electrolyte which could damage the electrolyte vessel.
As a rule of thumb: The sensor can be used for 12 months following manufacture and
storage.
Battery
The ULTRAMAT 23 must be operated continuously since the H
2
S sensor has its own battery
whose voltage is retained through operation of the ULTRAMAT 23.
With the analyzer switched off, the power supply to the sensor is from its own battery. The
resulting discharging of the battery impairs the sensor functionality and results in zero and
span drifts as well as increased signal noise This malfunction may last for 2 days or longer.
Materials used in the sample gas path
As a result of its polarity and good solubility in water, H
2
S accumulates on various materials.
These adsorption and desorption effects lead to increases in the response times. Therefore,
the inlet piping for the sample gas should be made of PTFE. Other materials should only be
used for short gas lines.
Ambient temperature
The influence of ambient temperature on the sensor is 3%/10 °C, referred to the full-scale
value; this corresponds to 1.5 vpm/10 °C.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...