Description
3.4 Technical specifications
ULTRAMAT 23
32
Manual, 3/2016, A5E37100388-003
Electrical characteristics
EMC interference immunity
(with safety extra-low voltage (SELV) with safe
isolation)
In accordance with standard requirements of NAMUR NE21 or
EN 61326-1
Power supply
100 V AC, +10%/-15%, 50 Hz,
120 V AC, +10%/-15%, 50 Hz,
200 V AC, +10%/-15%, 50 Hz,
230 V AC, +10%/-15%, 50 Hz,
100 V AC, +10%/-15%, 60 Hz,
120 V AC, +10%/-15%, 60 Hz,
230 V AC, +10%/-15%, 60 Hz
Power consumption
Approx. 60 VA depending on ordered version
Electrical inputs and outputs
Analog outputs
1 analog current output per component, 0/2/4/NAMUR ... 20 mA, float-
ing, max. load 750
Ω
Relay outputs
8, with changeover contacts, freely selectable, e.g. for fault, loading
capacity 24 V AC/DC/1 A, floating, non-sparking
Binary inputs
3, dimensioned for 24 V, floating
•
Pump
•
AUTOCAL
•
Synchronization
Serial interface
ELAN (RS485), PROFIBUS-PA/DP as option
AUTOCAL function
Automatic calibration with ambient air or nitrogen (depending on meas-
ured component),
adjustable cycle time from 0 (1) … 24 hours
Options
Add-on electronics, with 8 additional digital inputs and 8 additional relay
outputs, for e.g. triggering of automatic calibration, PROFIBUS PA/DP
Climatic conditions
Permissible ambient temperature
•
During operation
•
During transportation and storage
See specific technical specifications for IR detector/sensors (Infrared
detector (Page 35))
See specific technical specifications for IR detector/sensors (Infrared
Permissible ambient humidity
< 90% RH (relative humidity) during transportation and storage (Delivery
information (Page 14))
Permissible ambient pressure
See specific technical specifications for IR detector/sensors (Infrared
Maximum operating altitude
2000 m above sea level
Pollution degree
2
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...