Functions
8.2 Calibration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
115
8.2.3.2
Calibration: O2 measuring range: Calibrating the O2 zero point
You can use this function to re-adjust the zero point of the H
2
S
sensor with nitrogen. Connect nitrogen to the sensor and
commence the calibration with <ENTER>.
Following calling of the correction function, the current oxygen
value is displayed in the second line. If the displayed value
does not deviate by more than 1% from the set value, it is used
as the new zero point.
If the deviation is greater than 1% (as is the case in the exam-
ple on the left, see third line), a fixed default value is used in-
stead.
Note
The gas exchange takes place very slowly with low oxygen concentrations. In such cases we
recommend flow periods of approx. 30 minutes before you use the current value.
8.2.3.3
Calibration: O2 measuring range: Calibrate measuring range
You can calibrate the sensitivity of the electrochemical O
2
sensor using this function.
The adjacent menu display appears when you select the func-
tion.
Start the calibration procedure by positioning the cursor to the
3rd line and pressing the <ENTER> key.
The analyzer interrupts the current measurement when the
calibration procedure is selected and expects a flow of calibra-
tion gas.
The entered setpoint is displayed in the first line and the actu-
ally measured value in the second line.
If the analyzer recognizes a flow of calibration gas, the display
changes as shown. If the measured value in the second line
remains constant for more than approx. 10 s or does not
change significantly, press the <ENTER> key.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...