Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
153
Configuration: Device test: Inputs/outputs: Test of relays
You can use this function to test the status of the relays con-
trolled by this analyzer.
First enter the relay to be tested in this dialog. The analyzer can
control up to eight relays, or up to 16 relays with an add-on
board, whose contacts you can test. Connect an ohmmeter to
the corresponding relay outputs on the plug.
You can process the following parameters:
•
Relay number:
One of the relays 1 to 8 (relays 1 to 16 in the case of
analyzers with add-on board)
•
Condition:
The current state of the selected relay (active or inactive;
inactive in the Fig.).
The pin assignments of plug X80 (motherboard) und X50 (add-
on board) are described in section Pin assignments (Page 54).
You can exit the test by pressing the <ESC> key.
Configuration: Device test: Inputs/outputs: Test external solenoid valve
You can use this function to trigger external solenoid valves for
the zero and calibration gas supplies via the relay contacts.
Use the arrow keys <
↑
> and <
↓
> to select the zero gas valve
(ZV) in the first line or the calibration (span) gas valve (SV) in
the third line, and call the selected valve using the <ENTER>
key.
Switch the previously assigned relay using any arrow key in the
second or fourth line (the value on the right edge of the line
toggles between OFF and ON). The currently measured values
are output in the menu displayed during the test.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...