Functions
8.3 Parameter
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
125
8.3.1
Parameter: Measuring ranges
In this menu you can:
•
Permit or cancel the switching over between measuring
ranges
•
Set the full-scale values
•
Define a hysteresis.
Note
Please note that the range parameters only refer to the measuring ranges at the analog
outputs (see section Configuration: Inputs/outputs/pump: Analog outputs (Page 134)). The
display always shows the complete, physically possible range.
8.3.1.1
Parameters: Measuring ranges: Switch ranges
In the second line of this dialog you can set the measuring
range 1 or 2 or permit autoranging (automatic switching over
between these two ranges).
The 'Actual range' parameter can have the following values:
•
1:
The analyzer is set to the smaller range (MR 1).
•
2:
The analyzer is set to the larger range (MR 2).
•
1/2:
The analyzer is set to the larger range (MR 2). The start-of-
scale value of the analog output corresponds in this case to
the full-scale value of the smaller range (MR 1), the full-
scale value of the analog output corresponds to that of the
larger range (MR 2).
The result is that the analog output of the analyzer has a
range with zero offset (e.g. 90 ... 100%).
•
AR:
The analyzer switches over automatically from one range to
the other (AR = autoranging). Setting of the switchover
criteria is described in section Parameters: Measuring
ranges: Hysteresis (Page 127).
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...