Functions
8.4 Configuration
ULTRAMAT 23
154
Manual, 3/2016, A5E37100388-003
Configuration: Device test: Inputs/outputs: Test of inputs
You can use this function to test the status of the analyzer in-
puts. You can test the following inputs:
•
CAL, pump, SYNC (on the motherboard)
•
Binary inputs (on add-on board)
After calling this dialog, apply a voltage of 24 V DC to one of the
three inputs to be tested. The result is displayed in the fourth
line (here: "No active input").
The pin assignments of plug X80 (motherboard) und X50 (add-
on board) are described in section Pin assignments (Page 54).
Result of the CAL, pump, SYNC inputs test
Result of the binary inputs test
The adjacent error message occurs if an attempt is made to call
this function for a device without add-on board.
8.4.3.4
Configuration: Device test: Chopper/IR source
You can use this function to switch the chopper and IR source off for test purposes.
NOTICE
Device failure
Incorrect execution of this function may make the analyzer permanently incapable of
measuring!
Therefore this function must only be carried out by qualified servicing personnel.
Note
The analyzer is not ready for measurements for a certain period if the IR source or chopper
has been switched off. To reestablish the measuring capability, you must therefore provide a
sufficiently long warm-up phase depending on the switch-off period, e.g. by restarting the
analyzer.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...