Service and maintenance
10.2 Maintenance work
ULTRAMAT 23
170
Manual, 3/2016, A5E37100388-003
10.2.6
Maintenance work on the bench-top unit
10.2.6.1
Emptying the condensation trap
Proceed as follows:
1.
Switch off the pump by pressing the <PUMP> key.
2.
Disconnect the analyzer from the power supply.
3.
Loosen the condensation trap on the front of the analyzer by tilting it slightly, and pull
downwards carefully.
4.
Empty the trap and dispose of the condensate according to the composition of the sample
gas.
5.
Push the condensation trap on again from below.
CAUTION
Condensate in the analyzer
If condensate is spilled onto the analyzer during this process, it can penetrate into the
analyzer through gaps in the housing.
Such an analyzer is not suitable for measurements, and therefore must not be used any
longer!
10.2.6.2
Replacing the coarse filter
Proceed as follows for this:
1.
Switch off the pump by pressing the <PUMP> key.
2.
Loosen the condensation trap on the front of the analyzer as described in
section Emptying the condensation trap (Page 170).
3.
Remove the contaminated filter.
4.
Insert the new filter.
5.
Push the condensation trap on again from below.
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...