Service and maintenance
10.2 Maintenance work
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
171
10.2.7
Replacing the electrochemical oxygen sensor
WARNING
Danger of chemical burns
The O
2
S sensor contains acetic acid, which leads to burns on unprotected skin. When
replacing the sensor, its enclosure must not be damaged.
If contact with the acid occurs despite great care being taken, rinse the affected skin
immediately with plenty of water!
Also note that an exhausted or faulty O
2
sensor is hazardous waste and must be packed
and disposed of accordingly!
To replace the sensor, proceed as follows:
1.
Unscrew the two screws of the front cover and remove the cover.
2.
Unlock the plug of the sensor connection, and remove.
3.
Unscrew the O
2
sensor out of its holder.
4.
Remove the gasket of the O
2
sensor.
The exhausted O
2
sensor must be disposed of as electronic waste with the code number
160215 "Dangerous component removed from used equipment". It must therefore be
disposed of correctly and in an environmentally-friendly manner by a local waste disposal
company.
5.
Insert the new gasket.
6.
Screw in the new O
2
sensor and tighten hand-tight.
7.
Reconnect the plug.
8.
Enter the date of installation of the new sensor in the menu item "O
2
cal. after install" as
described in section Calibration: O2 measuring range: Sensor inst. date (Page 114).
9.
Calibrate the zero point of the new sensor as described in section Calibration: O2
Содержание ULTRAMAT 23
Страница 2: ......
Страница 10: ...Table of contents ULTRAMAT 23 10 Manual 3 2016 A5E37100388 003 Index 239 ...
Страница 18: ...Safety instructions 2 2 Analyzers in hazardous areas ULTRAMAT 23 18 Manual 3 2016 A5E37100388 003 ...
Страница 66: ...installation ULTRAMAT 23 66 Manual 3 2016 A5E37100388 003 ...
Страница 88: ...Commissioning 6 5 System setup with several analyzers in parallel ULTRAMAT 23 88 Manual 3 2016 A5E37100388 003 ...
Страница 147: ...Functions 8 4 Configuration ULTRAMAT 23 Manual 3 2016 A5E37100388 003 147 ...
Страница 160: ...Functions 8 5 Automatically executed functions ULTRAMAT 23 160 Manual 3 2016 A5E37100388 003 ...
Страница 164: ...Application note 9 1 H2S sensor with small measuring range ULTRAMAT 23 164 Manual 3 2016 A5E37100388 003 ...
Страница 184: ...Taking out of operation and disposal 12 2 Scrapping the analyzer ULTRAMAT 23 184 Manual 3 2016 A5E37100388 003 ...
Страница 194: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 194 Manual 3 2016 A5E37100388 003 ...
Страница 195: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 Manual 3 2016 A5E37100388 003 195 ...
Страница 196: ...Spare parts accessories 13 5 IR analyzer units ULTRAMAT 23 196 Manual 3 2016 A5E37100388 003 ...
Страница 229: ...Appendix A 4 Returned delivery ULTRAMAT 23 Manual 3 2016 A5E37100388 003 229 ...
Страница 230: ...Appendix ULTRAMAT 23 230 Manual 3 2016 A5E37100388 003 ...
Страница 238: ...List of abbreviations C 1 List of abbreviations ULTRAMAT 23 238 Manual 3 2016 A5E37100388 003 ...