System Overview
99
3.
Wait approximately 20 seconds.
4.
Perform the normal startup operation and open the Startup
window.
5.
At the Startup window, enter a password.
6.
Select
Re-Start
, then select
OK
.
7.
If possible, repeat the task prior to the power failure and verify that the data were
stored.
8.
If reagent was dispensed, you must perform a Weekly Wash before resuming
operation.
Additional Operating Instructions
To avoid errors or interruptions during operation, please read the following before using
the Chemistry System.
Switching Process Sequence and System Test List numbers
To avoid having quality control data reported for the wrong tests,
do not switch Process
Sequence Numbers
after control results have been accumulated. Patient results are also
affected.
To avoid incorrect reporting of results at the Review/Edit, the Daily QC windows, and the
Print Report form,
do not modify the order of the test items
at the System Test List
window after initial setup. If you reposition any test items at the System Test List window
after initial setup, results for samples already run could be associated with the wrong test
name.
Loading the reagent trays
CAUTION
Load each reagent container (wedge) on the correct reagent tray to avoid erroneous
results when using reagent containers without a barcode. Reagent Tray 1 (RTT-1)
is the front tray, while Reagent Tray 2 (RTT-2) is the rear tray.
The operator is alerted to this problem by the QC results.
Corrective action is to reload the reagents on the correct trays.
Updating the Cumulative QC
Update the QC Cumulative window before more than 200 daily control samples are run.
The QC Daily Precision Control window can manage a maximum of 200 results for each
control product (level). If an additional control sample is run, the current 200
th
result
(Px200) is overwritten by the new control result. Similarly, if multiple repetitions of a
control are requested, the system aspirates only one sample and its result replaces the
existing 200
th
result (Px200).
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...