
Maintenance
115
Daily maintenance
Inspecting and cleaning the probes
Materials required:
• Phillips screwdriver
• Alcohol prep pad or lint-free towels and 5% bleach
solution
Time: 10 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1.
Visually inspect each probe daily.
1
Dilution Probe (DPP)
2
Sample Probe (SPP)
3
Reagent Probe 1 (RPP1)
4
Reagent Probe 2 (RPP2)
Figure 5-1. Location of probes
•
Replace any clogged probes. See Replacing the Probes in the online Operator’s
Guide.
•
Clean all probes daily (proceed to step 3).
TIP
Perform the Shutdown Wash and Weekly Wash as scheduled, to prevent the probes
from clogging.
2.
Set the analyzer to Standby.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...