System Overview
75
Operating condition
Normal
indicator
Abnormal indicator and corrective
action
Diluent vol.
Indicates the status of the liquid volume of
diluent (saline solution). Insufficient liquid
is detected by a liquid level sensor in the
bottle.
OK
NG
Fill the diluent bottle with 0.9% saline
solution.
Make sure the sensor is in the proper
position.
Cuv. Wash Vol.
Status of the liquid volume of cuvette wash
solution. Insufficient liquid is detected by a
liquid level sensor in the bottle.
OK
NG
Fill the cuvette wash solution bottle.
Make sure the sensor is in the proper
position.
Cuv. Cond. Vol.
Status of the liquid volume of cuvette
conditioner. Insufficient liquid is detected
by a liquid level sensor in the bottle.
OK
NG
Fill the cuvette conditioner bottle. Make
sure the sensor is in the proper position.
Conc.wast.tank
(if installed)
Status of the volume of the optional
concentrated waste tank. A sensor monitors
whether the tank is filled with waste.
OK
NG
Empty the concentrated waste tank.
If there is still a problem, check the tubes
for obstructions.
Waste tank
Status of the volume of the optional waste
tank. A sensor monitors whether the tank is
filled with waste.
OK
NG
Verify that the waste tubes are free of
obstructions. If the problem persists,
contact your local support provider or
distributor.
3.
Check the availability of the reagents and wash solutions:
WARNINGS
•
Read and follow the cautions on the box and label before handling any reagents.
Some method and system reagents are classified as hazardous material.
•
Avoid contact of method or system reagents with skin and eyes. These materials
can cause infection and burns. Wear suitable protective clothing, gloves, and
eye/face protection. In case of contact with eyes, rinse immediately with plenty
of water and seek medical advice. In case of contact with skin, immediately
wash with soap and water.
•
Avoid inhaling chemical vapors. If you inhale chemical vapors, promptly leave
the area and seek fresh air.
CAUTIONS
•
Turn the analyzer to
Standby
or turn the main power switch to
OFF
if any
water or reagent spills into the analyzer to avoid injury or damage to the system.
Using alcohol preps or lint-free towels moistened with 5% bleach solution,
clean the surfaces and trays (see
Maintenance
section for detailed procedures).
If any liquids spilled inside the analyzer, contact your local technical support
provider or distributor.
•
Use only Siemens-recommended ancillary reagents to avoid erroneous test
results.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...