56
ADVIA 1800 Operator’s Guide
2. To reactivate a reagent container, select the Deselect checkbox again.
Printing a Reagent Summary Report
1. Select Print RGT Summary.
2. When the Print Reagent Summary? message box opens, select Yes to print the report.
Three reports print:
•
a summary report for RTT1
•
a summary report for RTT2
•
a Total Tests summary report
Viewing Total Test Summary
1. Select Total Test Summary.
The Total Test Summary window displays a list of all methods on-board and their total
tests counts. The Total Test Summary window only includes paired reagent packs, and
packs that have a 'Passed' status for the most recent calibration. If there is only 1 R1 and 1
R2 reagent present, the total tests value is the lower of the tests remaining in the R1 and
R2 reagent container. If multiple R1 and R2 reagents are present on the system, the value
displayed is the minimum of the sum of the combined R1 reagents and the sum of the
combined R2 reagents. For example:
•
R1a: 20 tests + R1b: 5 tests + R1c: 10 tests = 35 remaining
•
R2a: 5 tests + R2b: 10 tests + R2c: 10 tests = 25 remaining
•
Total Tests = 20 tests remaining
2. To print the Total Test Summary, select Print.
3. To close the Total Test Summary, select Exit.
Viewing Cal Interval Review
Use the Cal Interval Review list to review daily which methods you should calibrate.
NOTE
For barcoded reagents, this information is preset to Siemens guidelines. If this field
displays a 0, no Cal Interval Days value shows in the Reagent Inventory window. The
operator has an option to modify this field.
•
The calibration is reset to its default value only if a full calibration is performed. If an
RBL is performed, the Cal Interval Days column is not reset to its default value.
1. Select Cal Interval Review.
•
The CalInterval Reviewwindowopens. Insteadofanalphabeticallist,thiswindowre-
sortsthe Reagent Inventory list methods in order of Cal Interval Days remaining.
Methods with the shortest days remaining are at the top.
2. To print Cal Interval Review, select Print.
3. To close the Cal Interval Review, select Exit.
CTT Monitor window
Use this window to monitor the amount of liquid in containers on the CTT tray.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...