ADVIA 1800 Operator’s Guide
WARNING
To prevent infection from contacting serum directly, wear suitable protective
gloves when you remove the electrode from the solution.
NOTE:
High-concentrated salt water is used as a preservation solution to maintain
electrode performance. When the electrode package is opened, wash the electrode
with sufficient water and wipe well before use. Small amounts of salt on the electrode
may cause rust on the electrode connector.
NOTE:
To store the reference electrode, refer to Storing the Reference Electrode.
8.
Replace the electrodes on the instrument with the newly conditioned ones.
9.
Calibration is performed as part of the electrode replacement.
10.
If the calibration fails, repeat the calibration.
Cleaning the dilution bowl and waste-drain nozzle
Materials required:
•
Cotton stick
•
Deionized water
•
Household bleach
•
Philips screwdriver
Time: 45 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Cleaning the dilution bowl
1.
At the Menu Panel, select
Maint.
, then select
ISE Operation
.
2.
At the ISE Operation window, in the Period.wash area, select
OFF
, then select
Set
.
3.
Using a Phillips screwdriver, remove the screws (
1
, see Figure 5-2) that secure the
DPP shield to the analyzer panel.
4.
Push the DPP shield to the right and slowly lift the DPP shield until it reaches
approximately a 90° angle, then gently lift the tab of the DPP shield and remove.
5.
Remove the Phillips screws, then remove the ISE cover.
6.
Loosen the screw retaining the stainless steel cover at the top of the ISE unit, and
remove that cover by sliding it toward you.
7.
At the ISE Operation window, next to Final operation, type
16
in the field next to
Pure water position.
8.
Select container 1 setting for 10-mL tube.
9.
Fill a 10-mL tube with deionized water and place it on the CTT tray in position 16.
10.
At the ISE Operation window, in the Final operation area, select
Execute
.
Water is dispensed into the ISE module.
11.
To dissolve the crystals attached to the liquid-supply nozzle, let it stand for about five
minutes.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...