Maintenance
135
12.
Only
if you replaced the lamp, select
Regist Data
, then select
OK
in the Registration
window.
13.
If not, proceed to step 14.
NOTE:
The system uses the data from the lamp energy data registration as the
comparison standard for the next calculation of the attenuation ratio. The displayed
attenuations for all the reference values are set to 100%, indicating no attenuation.
14.
At the Menu Panel, select
System(s)
, then select
Screen Print
to print the window
contents.
15.
Save the printout with your laboratory maintenance records.
16.
Exit the Lamp Energy Monitor window, then select
Initialize
to switch the system
from the WAIT
state to the READY
state.
17.
Run cell blank measurement.
Reading lamp energy data
1.
Execute the lamp energy check and acquire the data.
•
A graph displays at the window.
•
The lamp energy check data displays to the left of the graph.
2.
Select a wavelength for which to display data.
3.
Display the voltage or the AD value.
4.
(Optional) Change the vertical scale of the graph.
Measuring cuvette blanks
NOTE
: When performing weekly maintenance, be sure to perform the procedures in the
following sequence:
1.
Weekly wash
2.
Lamp energy check
3.
Cuvette blank measurement (this procedure)
Materials required:
• No materials required
Time: 20 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Reaction cuvettes cause changes in absorption with use. After the weekly wash, perform
the cuvette blank measurement to determine the change. The cuvette blank is only run
weekly, even if your lab runs the Weekly Wash as a daily procedure.
1.
At the Menu Panel, select
Maint
, then select
User Maintenance
.
2.
In the Cell blank meas. check
area, select
Start CB
.
The measured cuvette blank values for all RRV cuvettes and a list of abnormal cells are
available in the real-time monitor in approximately 15 minutes.
3.
Select
Yes
to save the data.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...