Maintenance
149
Every 3 months maintenance
Replacing the lamp
Materials required:
• Halogen lamp, 12 V/50 W
(REF 02127928, PN 073-0099-01)
Time: 60 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
The lamp MUST be replaced:
•
Quarterly
•
After approximately 2000 hours of use
•
If the system warns that lamp energy is out of range
•
If the Checking Lamp Energy procedure indicates the A-D points are outside the ±40
range of the scatter plot center line or the volts for any of the wavelengths are outside
the range of 5.0 to 9.0 volts
•
If the ATTENU(%) for any of the 14 wavelengths on the Lamp Check Energy
window falls below 80%
1
Lead Wire Connector
2
Lead Wire Connector
3
Lamp Screws
4
Lamp Plate
5
Alignment Hole and Pin
Figure 5-22. Components of the lamp
1.
Set the analyzer to
Standby
.
2.
Lift and remove the access panel in front of the Rotating Reaction Tray to expose the
lamp housing.
WARNING
The lamp housing is
hot
. To avoid burns, allow it to cool down
(approximately 10 minutes) before touching any components.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...