Maintenance
119
11.
To remove the probe cleaning tool, hold the joint-
holder end of the probe and carefully turn the push-
in fitting counter clockwise.
12.
Reinstall the probe.
13.
Turn the Standby Operate Switch to Operate and
reconnect.
14.
To ensure proper flow through the probe, at the
Operation Panel, perform a PRIME 2.
NOTE:
To remove the tubing from the push-in fitting, press the release ring towards the
fitting and pull the tubing out. No tools are required. To install the tubing, simply push the
tubing in.
Inspecting and cleaning the mixing rods and mixer wash cups
Materials required:
• Lint-free towel
• Deionized water
• Cotton-tipped applicators
Time: 5 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Figure 5-4. Mixing rods
1
Dilution Mixer
2
Reagent Mixer1
3
Reagent Mixer2
1.
Visually inspect each mixing rod and mixer wash cup for cleanliness.
2.
Clean any dirty mixing rods or wash cups to avoid contamination of the mixers, which
results in carryover:
a.
Set the analyzer to Standby.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...