Clinitek Status Connect System
48
Operator’s Guide
Connecting to the LAN
To connect the system to the LAN, perform the following steps:
1. At the
Select Ready
screen, select
Instrument Set Up
.
The
Choose Settings
screen displays.
2. Use the arrow keys to select
Instrument Settings
.
3. Select
Select
.
The
Instrument Settings
screen displays.
4. Use the arrow keys to select
Connectivity
.
5. Select
Select
.
The
Connectivity
screen 1 of 2 displays.
6. To enable communication between the instrument and the connector,
select
Enabled
.
To prevent communication, select
Disabled
.
7. To automatically send new and recalled patient results to the LIS or a PC
directly connected to the system, select
Enabled
.
To prevent sending new and recalled patient results to the LIS or PC,
select
Disabled
.
8. Select
Next
.
The
Connectivity
screen 2 of 2 displays.
9. Use one of the following four options for network connectivity.
•
To connect to the LAN using the Ethernet cable, select
Wired
connection
, select
Edit wired settings
, and see below.
•
To connect to the LAN using the wireless card, select
Wireless
connection
, select
Edit wireless settings
, and see below.
•
To connect to an external computer system using the serial
connection, select
Serial connection
and refer to the
Clinitek Status
Analyzer Operator’s Manual
.
•
For no connectivity, select
None
.
10. Select
Done
3 times to return to the
Select Ready
screen.
NOTE:
If there is no connector, Siemens Healthcare Diagnostics recommends
setting the connector to Disabled setting. Setting the connector to Enabled
setting without a connector may prevent communication with an external
system.
NOTE:
You can edit connectivity settings if the connector is disabled.
Connecting to the LAN, Wired Settings
Before beginning, ensure the system is connected to the LAN with the
supplied Ethernet cable.
Refer to
Connecting to the LAN
above.
Содержание Clinitek Status Connect System
Страница 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Страница 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Страница 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Страница 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Страница 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Страница 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Страница 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Страница 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Страница 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Страница 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Страница 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Страница 72: ...Clinitek Status Connect System 72 Operator s Guide ...