![Siemens Healthcare Clinitek Status Connect System Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/siemens-healthcare/clinitek-status-connect-system/clinitek-status-connect-system_operators-manual_1266125030.webp)
Clinitek Status Connect System
30
Operator’s Guide
4. To have the instrument determine QC pass or fail, select
Instrument
.
To have the operator determine QC pass or fail, select
Operator
.
5. To permit patient tests if the QC test fails, select
Yes
.
To prevent patient tests if the QC test fails, select
No
.
6. Select
Next
.
The
QC Settings-Cassette test
screen 3 of 3 displays.
NOTE:
If you selected to have the operator determine pass or fail in Step 4
above,
Defining QC Cassette Test Controls
is not necessary.
Defining QC Cassette Test Controls
To define QC cassette test controls, perform the following steps:
1. To set the number of levels to complete a QC test, select
1
or
2
.
2. For each control set in Step 1, select the corresponding
Control Level, 1
or
2
.
For example, if you set 2 controls, you must define Control Level 1 and
Control Level 2.
The
Control Level 1-Cassette test
screen displays.
3. Select
Enter name of control
.
The
Name of Control
screen displays.
4. Enter the name of this control.
Use the alpha keyboard to enter text.
To enter numeric text, select
123
.
5. Select
Enter
.
The
Control Level 1-Cassette test
screen displays.
6. To set the control level as positive, select
Positive
.
To set the control level as negative, select
Negative
.
7. Select
Done
.
The
QC Settings-Cassette test
screen 3 of 3 displays.
8. If necessary, repeat defining QC cassette test controls for the next control
level.
9. When all QC levels are defined, select
Done
.
The
QC Settings
screen displays.
Setting QC Cassette Schedule
To set the QC cassette test schedule, perform the following steps:
1. For QC cassette test, select
Interval
.
The
QC Settings-Set QC schedule for cassette test
screen displays.
2. To schedule QC tests by hours, select the
Hours
radio button and use the
arrow keys to indicate the number of hours between QC tests.
Go to Step 7.
To schedule QC tests by days, select the
Days
radio button and use the
arrow keys to indicate the number of days between QC tests.
Содержание Clinitek Status Connect System
Страница 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Страница 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Страница 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Страница 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Страница 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Страница 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Страница 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Страница 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Страница 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Страница 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Страница 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Страница 72: ...Clinitek Status Connect System 72 Operator s Guide ...