Operator’s Guide
Cause Corrective
Action
Deterioration of the quality
control product.
Prepare control solution using a fresh
bottle of control product.
Repeat the quality control procedure.
If the fresh control solution fails to give
results within the expected values,
proceed to the next possible cause.
Clinitek Status analyzer
malfunction.
If you cannot successfully complete the
quality control procedure, an analyzer
malfunction or reagent strip problem may
exist. Check the calibration bar for dust,
dirt, or scratch marks, or contact your
local technical support provider for
assistance.
Содержание Clinitek Status Connect System
Страница 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Страница 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Страница 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Страница 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Страница 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Страница 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Страница 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Страница 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Страница 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Страница 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Страница 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Страница 72: ...Clinitek Status Connect System 72 Operator s Guide ...