![Siemens Healthcare Clinitek Status Connect System Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/siemens-healthcare/clinitek-status-connect-system/clinitek-status-connect-system_operators-manual_1266125031.webp)
Clinitek Status Connect System
31
Operator’s Guide
3. If you selected Days, select
Set QC time
.
The
QC Settings-Set QC time for cassette test
screen displays.
4. Select the
1
,
2
, or
3
radio button for the number of QC tests per day.
5. Use the arrow keys to indicate time of day for each test.
6. Select
Set
.
The
QC Settings-Set QC schedule for cassette tests
screen displays.
7. Select
Done
.
The
QC Settings
screen displays.
8. Select
Done
3 times to return to the
Select Ready
screen.
Running a QC Strip Test
NOTE:
Prepare the controls according to the manufacturer’s instructions.
This section describes how to run a QC strip test when it is due and for one
QC level. Repeat the procedure if more than one QC level is set.
To perform a QC strip test when due, perform the following steps:
1. At the
Select Ready
screen, select
QC Test Strip due
.
The
QC Test
screen displays.
2. Select
QC Strip Test Required
.
3. If the instrument is set to determine pass/fail, the
Control Lot
screen
displays. Go to Step 8.
If the operator is set to determine pass/fail, the
Name of Control
screen
displays.
4. Enter the control name.
Use the alpha keyboard to enter text.
To enter numeric text, select
123
.
5. Select
Enter
.
The
Name of Control Level
screen displays.
6. Enter the control level.
Use the alpha keyboard to enter text.
To enter numeric text, select
123
.
7. Select
Enter
.
The
Control Lot
screen displays.
8. Enter the control lot.
Use the alpha keyboard to enter text.
To enter numeric text, select
123
.
9. Select
Enter
.
The
Control Expiration
screen displays.
10. Use the arrow keys to indicate the control lot expiration date.
11. Select
Enter
.
The
Strip Lot
screen displays.
12. Use the alpha keyboard to enter text.
To enter numeric text, select 123.
13. Select
Enter
.
The
Strip Expiration
screen displays.
Содержание Clinitek Status Connect System
Страница 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Страница 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Страница 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Страница 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Страница 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Страница 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Страница 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Страница 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Страница 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Страница 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Страница 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Страница 72: ...Clinitek Status Connect System 72 Operator s Guide ...