Setting QC Strip Test Pass/Fail Ranges
To set QC strip test pass/fail ranges, perform the following steps:
1. Select
Set pass ranges
.
The
QC Level 1-Tests and pass ranges
screen 1 of 3 displays.
2. To set the glucose pass range, select the
GLU
radio button.
3. Select the corresponding
Pass range
button.
The
QC Level 1-Set pass range for GLU
screen displays.
4. Use the arrow keys to indicate the lowest acceptable clinical values for
glucose.
5. Use the arrow keys to indicate the highest acceptable clinical values for
glucose.
6. Select
Set
.
The
QC Level 1-Tests and pass ranges
screen 1 of 3 displays.
7. Enter the pass ranges for glucose, ketone, blood, bilirubin, specific gravity,
pH, protein, nitrite, albumin, urobilinogen, leukocytes, and creatinine as
directed in Steps 2–6.
Select
Next
to advance through all 3 screens.
NOTE:
Each test is included in the QC if the corresponding button is filled.
Each test is excluded if the corresponding button is blank. If you do not select
a radio button, that parameter is disabled for the QC test.
When you complete entering pass ranges for the tests, the
QC Level 1-
Tests and pass ranges
screen 3 of 3 displays.
8. Select
Done
.
The
Control Level 1-Strip test
screen displays.
9. Select Done.
The
QC Settings-Strip test
screen 3 of 3 displays.
10. If necessary, repeat defining QC strip test controls for the next control
level.
11. When all QC levels are defined, select
Done
.
The
QC Settings
screen displays.
Setting QC Strip Schedule
To set the QC strip test schedule, perform the following steps:
1. For QC strip test, select
Interval
.
The
QC Settings-Set QC schedule for strip tests
screen displays.
2. To schedule QC tests by hours, select the
Hours
radio button and use the
arrow keys to indicate the number of hours between QC tests.
Go to Step 7.
To schedule QC tests by days, select the
Days
radio button and use the
arrow keys to indicate the number of days between QC tests.
3. If you selected Days, select
Set QC time
.
The
QC Settings-Set QC time for strip tests
screen displays.
Содержание Clinitek Status Connect System
Страница 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Страница 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Страница 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Страница 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Страница 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Страница 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Страница 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Страница 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Страница 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Страница 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Страница 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Страница 72: ...Clinitek Status Connect System 72 Operator s Guide ...