Clinitek Status Connect System
29
Operator’s Guide
4. Select the
1
,
2
, or
3
radio button for the number of QC tests per day.
5. Use the arrow keys to indicate time of day for each test.
6. Select
Set
.
The
QC Settings-Set QC schedule for strip test
screen displays.
7. Select
Done
.
The
QC Settings
screen displays.
8. Select
Done
3 times to return to the
Select Ready
screen.
Setting Up QC for Cassette Tests
This section describes how to set QC for hCG cassette tests and define their
type, frequency, and pass/fail ranges. You can set required or reminder QC
tests, frequency and schedule, and the number of control levels for each QC.
Using the QC Set Up features, you can set whether the instrument or operator
determines QC pass or fail results. If the instrument determines pass/fail, it
compares the QC results to ranges you specified in the QC Set Up procedure.
Or, you can have the operator determine QC test pass/fail after viewing the
results.
Accessing QC Cassette Set Up
To access the QC Set Up procedure for cassette tests, perform the following
steps:
1. At the
Select Ready
screen, select
Instrument Set Up
.
The
Choose Settings
screen displays.
2. Use the arrow keys to select
Instrument Settings
.
3. Select
Select
.
The
Instrument Settings
screen displays.
4. Use the arrow keys to select
QC Settings
.
5. Select
Select
.
The
QC Settings
screen displays.
6. For QC cassette test, select
Set Up
.
The
QC Settings-Cassette test
screen 1 of 3 displays.
Setting QC Cassette Tests
To set QC cassette tests, perform the following steps:
1. To allow QC prompts, select
Enabled
.
To prevent QC prompts, select
Disabled
.
2. To require QC tests when they are due and not permit patient tests, select
Required
.
To permit patient tests even if a QC test is due, select
Reminder
.
3. Select
Next
.
The
QC Settings-Cassette test
screen 2 of 3 displays.
Содержание Clinitek Status Connect System
Страница 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Страница 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Страница 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Страница 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Страница 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Страница 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Страница 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Страница 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Страница 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Страница 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Страница 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Страница 72: ...Clinitek Status Connect System 72 Operator s Guide ...