12. Statim Data Logger Troubleshooting
This troubleshooting addresses problems that may be encountered when operating the
Data Logger. In this troubleshooting, MSD (Mass Storage Device) refers to both USB
memory sticks and SD Flash cards. The headings, such as, “SAFELY REMOVE MSD <>
MSD/FLASH DETECTED”, refer to the display of multiple messages in alternation of one
another on the Stat
im
screen.
MSD/FLASH = Mass Storage Device/Flash Card. ie; SD Memory Card or USB Flash Drive
T1. If display reads:
MSD NOT CONNECTED <> INSERT MSD/FLASH
Follow the steps below to ensure everything is setup properly.
1. Check the serial cable connection.
2. Check the power connection.
3. Ensure the lower red LED is lit. It is located between the Data Logger’s serial port
and power input.
4. Check that the Mass Storage Device (MSD) is properly inserted in the
appropriate slot of the Data Logger.
5. Repeat the instructions for “Installing the SciCan Data Logger on your Statim”.
T2. If display reads:
SAFELY REMOVE MSD <> MSD/FLASH DETECTED
The MSD/Flash can be safely removed without affecting the data.
T3 If display reads:
MSD/FLASH FULL <> REPLACE MSD
The MSD is full. Export the data.
T4. If MSD has:
Missing lines of Data on MSD/FLASH
Refer to the STATIM screen to confirm successful sterilization.
Reset the Data Logger by unplugging its power, disconnecting the MSD
and waiting 10 seconds. Then re-connect the power adapter and insert
the MSD into the Data Logger.
If the problem persists, contact the SciCan Service Center.