11. Statim 2000S and 5000S Troubleshooting
Cycle Faults – Software Revision R4xx, R5xx and R6xx
52
Cycle Fault #55
For the Rubber and Plastics cycle, the theoretical
cassette temperature calculated from the measured
cassette pressure was more than 4°C above the
sterilization temperature, allowing for measurement
error (i.e. Pchm > 232 kPa).
Pressure and/or temperature sensor miscalibration.
Unable to depressurize cassette (blocked exhaust,
failed valve).
Cycle Fault #60
For a 134°C cycle, the chamber temperature
dropped below the sterilization temperature, allowing
for measurement error (i.e. Tchm < 134°C).
Chamber temperature sensor miscalibration. Poor air
removal during conditioning (exhaust blockage). Not
able to generate steam or a leak in the system
Cycle Fault #61
For a 134°C cycle, the chamber temperature rose
more than 4 C° above the sterilization temperature,
allowing for measurement error (i.e. Tchm > 138°C).
Chamber temperature sensor miscalibration. Unable
to depressurize cassette (blocked exhaust, failed
solenoid valve).
Cycle Fault #62
For a 134°C cycle, the theoretical chamber
temperature as calculated from the measured
chamber pressure was more than 2°C below the
measured chamber temperature, allowing for
measurement error.
Pressure and/or temperature sensor miscalibration.
Unable to depressurize cassette (blocked exhaust,
failed valve). Poor air removal (partial exhaust
blockage).
Cycle Fault #63
For a 134°C cycle, the theoretical chamber
temperature as calculated from the measured
chamber pressure was more than 2°C above the
measured chamber temperature, allowing for
measurement error.
Pressure and/or temperature sensor miscalibration.
Unable to depressurize cassette (blocked exhaust,
failed valve). Poor air removal (partial exhaust
blockage).
Cycle Fault #64
For a 134°C the theoretical cassette temperature
calculated from the measured cassette pressure was
below the sterilization temperature, allowing for
measurement error (i.e. Pchm < 304 kPa).
Pressure and/or temperature sensor miscalibration.
Unable to generate steam, or there is a leak in the
system.
Cycle Fault #65
For a 134°C cycle, the theoretical cassette
temperature calculated from the measured cassette
pressure was more than 4°C above the sterilization
temperature, allowing for measurement error (i.e.
Pchm > 341.2 kPa).
Pressure and/or temperature sensor miscalibration.
Unable to depressurize cassette (blocked exhaust,
failed valve).