10. Statim 5000 Trouble-shooting – Cycle Faults
44
Cycle Fault #13
N/A
Cycle Fault #14
Software Revisions
R1xx or R2xx only
The steam generator temperature is above 171°C
during the sterilization phase of a cycle.
This may be the result of a blocked steam generator
inlet tube, a failed or weak water pump, a failed pump or
steam generator triac, faulty seal installation or failed
solenoid valve.
Cycle Fault #15
The cassette temperature raised above the high
threshold during the Sterilization phase of a cycle or
above 138.6°C during conditioning or pressurizing
phase of the cycle
Probable causes are a blocked duct in the cassette, a
pinched exhaust tube leading to the waste bottle, or a
faulty solenoid valve.
Cycle Fault #16
The steam generator temperature went above a
threshold value.
This may be the result of a blocked steam generator
inlet tube, a failed or weak water pump, a failed pump
triac, failed steam generator triac, faulty seal installation
or failed solenoid valve.
Cycle Fault #17-
#18
N/A
Cycle Fault #19
For software rev. R1xx, R2xx and R4xx the steam
generator calibration is invalid. For software rev.
R5xx and R6xx the validation thermocouple
calibration is invalid. In both cases a new steam
generator or validation thermocouple calibration is
required.
This occurs when a new controller board or
microprocessor has been installed. This may also
happen when the unit has been subjected to a strong
static discharge corrupting the memory. Calibrate the
boiler / validation thermocouples
Cycle Fault #20
Software Revisions
R1xx or R2xx
The pump has failed to pump water into the steam
generator during a pre-vent pump time-out. The
steam generator temperature was greater than
140°C for 3.6 seconds after the pump was activated
to pump water to cool the steam generator.
This may be the result of a blocked steam generator
inlet tube, a failed or weak water pump, a failed pump
triac or steam generator triac, faulty seal installation or
failed solenoid valve. Also may be the result of a
defective water quality sensor (insufficient water in the
reservoir)
Cycle Fault #21-
#24
N/A