11. Statim 2000 Trouble-shooting
Printer Messages
54
Cycle Fault Number
Description of Fault
Probable Cause of Fault
Message "PRINTER
FAULT" (If optional printer
is installed)
Printer is not printing.
This may be caused by a paper
jam, a failed printer or failed printer
electronics. Check that power is
ON, check connector attachments.
"CYCLE ABORTED"
This error message is displayed on the printout only,
followed by the message "NOT STERILE", as a result of the
operator pressing the STOP button to stop the cycle or as a
result of any other abnormal cycle termination, including
CYCLE FAULT errors.
"STOP BUTTON
PRESSED"
The operator pressed the STOP button to stop the cycle.
The LCD shows the message "NOT STERILE" as a result.
"CYCLE
INTERRUPTED"
This error is displayed on power-up following a power failure
occurs during a cycle or whenever the power is turned OFF
after an error occurred without pressing the STOP button to
reset. (Rev. R200 level code)
This message is displayed when the sterilization phase has
failed to start within three minutes of the cassette reaching
the sterilization temperature. If it occurs in three consecutive
cycles Cycle Fault #26 is displayed. (Software Rev. R4xx,
R5xx and R6xx)
"PRESS STOP TO
RESET"
This message is displayed for all error faults. The user
MUST press the STOP button on the keypad to reset the
unit; otherwise the user will be unable to initiate another
cycle.