2.1. Calibration
Calibrate thermocouple offsets
2.1.1. Non S units
Chamber Calibration Screen
VVV.V VO CCC.C CO
VC
Chamber Calibration Sterilization phase
VVV.V VO CCC.C CO
ADJUST TEMP! VC
“ADJUST TEMP!” is displayed during moments when the Chamber Offset should be
adjusted. Chamber Offset (CO) flashes when it can be adjusted.
Validation thermocouple validation
VVV.V VO* CCC.C CO
VC
Screen Representation
VVV.V
Validation thermocouple reading
VO
Validation thermocouple offset
CCC.C
Chamber thermocouple reading
CO
Chamber thermocouple offset
VC
Difference between Validation thermocouple reading and
Chamber thermocouple reading. After calibration, this value
should be close to 0.
Keypad functions when calibration cycle is not started:
Unwrapped
To start the Chamber thermocouple Calibration
cycle, first press and release the Unwrapped key
followed by the pressing of the Start key.
To start the Validation thermocouple Calibration
cycle, select the Unwrapped key together with the
Start key.
900/Sterimaster type units – Start the Validation
thermocouple Calibration cycle.
Wrapped
Send the last printout is to the printer. The printer
must be connected and selected for this feature to
work.
900 type units – Start the Chamber thermocouple
Calibration cycle.