29
2.14. Validation Offset
View the offset of the thermocouple
>Validation Offset
##
Screen Representation
##
Offset Value
Keypad
Rubber and Plastics
Return to main menu
Stop
Exit to normal mode of operation
2.15. Voltage Calibration
Adjust voltage offsets
Voltage Calibration
V = VVV VCal. = CCC
Screen Representation
VVV
Voltage measured by unit
CCC
Voltage calibration offset. This should be adjusted so that
the VVV value is the same as the line voltage measured by
a voltmeter.
Keypad
Unwrapped
Increase current field
Wrapped
Decrease current field
Rubber and Plastics
Select and return to main menu
Stop
Exit, without saving, to normal mode of operation
2.16. Voltage setup
Select input voltage of unit
>Voltage setup
115V
230V
Keypad
Unwrapped
Select next option. Second line shows the new value
Wrapped
Select previous option. Second line shows the new value
Rubber and Plastics
Select and return to main menu
Stop
Exit, without saving, to normal mode of operation