25
cycles and cycle parameters are displayed one by one on the LCD display. The
available sets of cycle parameters are listed in the chart below.
Available Cycles
Unlocked
mode
Locked mode
Keypad
Button
SOLID UNWRAPPED (N)
134
°
C / 3.5 min
HOLLOW UNWRAPPED (S)
134
°
C / 3.5 min
HOLLOW UNWRAPPED (S)
134
°
C / 18 min
Unwrapped
HOLLOW WRAPPED (S)
134
°
C / 3.5 min
HOLLOW WRAPPED (S)
134
°
C / 18 min
Wrapped
RUBBER / PLASTICS (S)
121
°
C / 15 min
RUBBER / PLASTICS (S)
121
°
C / 30 min
Rubber &
Plastics
AIR DRYING ONLY
60 MIN. MAXIMUM
Air Drying
2.4. Conductivity Setup
Adjust the water conductivity thresholds
CD=xx.xuS/NNN/y.yppm
L=LL.L H=HH.H G=G.GG
Screen Representation
xx.x
Water conductivity in uS (micro-Siemens)
NNN
Water conductivity in ADC (Analog to Digital converter)
counts (0…255)
y.y
Water quality in ppm (parts per million)
LL.L
Lower value threshold / No water threshold. Values lower
than this triggers “No Water” error (default 0.5uS)
HH.H
High value threshold / Water quality threshold. Values
larger than this triggers “No Water” error (default 10.0uS)
G.GG
Water conductivity circuit gain (default 1.00)
Note: Distilled water readings should be between Low and High thresholds.
The currently selected field is indicated by the flashing value on the display.