23
Rubber and Plastics
Start the Burn-in cycle
Stop
Exit to normal mode of operation
Keypad functions when calibration cycle is running:
Unwrapped
Increment selected offset
Wrapped
Decrement selected offset
Stop
Interrupt calibration cycle
2.1.2. S units
Chamber Calibration Screen
PPP.P PO HHH.H CO
CCC.C
Chamber Calibration, Pressure calibration phase
PPP.P PO HHH.H CO
Adjust PRESS. CCC.C
“Adjust PRESS!” is displayed during moments when the Pressure Offset at
Atmospheric (POA) should be adjusted. PO flashes when it can be adjusted. When the
phase ends, “Adjust PRESS!” disappears from the screen. CO will then start flashing,
showing that the Chamber Offset (CO) can be adjusted.
Chamber Calibration Sterilization phase
PPP.P PO HHH.H CO
ADJUST TEMP! CCC.C
“ADJUST TEMP!” is displayed during moments when the Chamber Offset (CO) should
be adjusted. CO flashes when it can be adjusted.
While in the sterilization phase of the calibration cycle, the Pressure Offset at
Sterilization (POS) can be adjusted. First, select PO using the keypad. Then increment
or decrement PO to match the external pressure reference. PO will flash when it is
selected and can be adjusted.
PPP.P
Pressure reading
PO
Pressure offset. It reads POA when the pressure is lower
than 115kPa and POS when the pressure reading is higher
or equal to 115kPa.
HHH.H
Calculated chamber pressure
CO
Chamber offset
CCC.C
Chamber temperature reading
Keypad functions when calibration cycle is not started:
Unwrapped