9. Statim 2000 Trouble-shooting
Trouble-shooting, Original Statim Error Codes Unit Displays Select A Program
41
#5
between 2 minutes 44 seconds and 2 minutes 24
seconds remaining in the cycle. If a request to pump
water occurs outside of 2 minutes 44 seconds,
Check Cassette #5 occurs.
a contaminated steam generator or a weak pump.
Check Cassette
#6
The steam generator temperature is more than 5°C
higher than the chamber temperature.
Most probable causes are a blocked duct, a blocked
solenoid valve, miscalibration or a pinched exhaust tube.
Check Cassette
#7
The cassette temperature has dropped below
130.5°C during the sterilizing phase of the
UNWRAPPED or WRAPPED Cycle, or below
117.5°C during the sterilizing phase of the RUBBER
AND PLASTIC Cycle.
Most probable cause is a faulty solenoid valve.
Check Cassette
#8
The chamber temperature is more than 5°C higher
than the steam generator temperature.
Most probable causes are thermocouple miscalibration or
very strong pump delivery.
Check Cassette
#9
This is a special error message for the U.K. market.
The chamber temperature has exceeded 139°C
during the UNWRAPPED/ WRAPPED Cycle, or has
exceeded 126°C during the RUBBER AND
PLASTIC Cycle.
Most probable causes are a blocked or faulty solenoid
valve or a pinched exhaust tube.
Service Needed
The steam generator temperature has exceeded
170°C, chamber temperature has exceeded 147°C,
or the thermocouple is faulty.
Most probable causes are a blocked or faulty solenoid
valve, a pinched exhaust tube, a broken thermocouple
lead, an open, disconnected or defective thermocouple, a
weak pump or a contaminated steam generator.
Check Cassette
(without a
number)
The cassette temperature has failed to drop to
103°C within a timeout period at the end of a cycle.
Most probable causes are a blocked exhaust duct in the
cassette, a blocked or faulty solenoid valve or a pinched /
kinked exhaust tube.