10. Statim 5000 Trouble-shooting – Cycle Faults
46
Cycle Fault #90
Software Revisions
R6xx only
Corrupted or not initialized chamber calibration
value
This occurs when a new controller board or
microprocessor has been installed. This may also
happen when the unit has been subjected to a strong
static discharge corrupting the memory. Calibrate
chamber thermocouple.
Cycle Fault #98
Software Revisions
R4xx and up
(includes R5xx and
R6xx)
Microcontroller fails to communicate with ADC
(Analog to Digital Converter)
Hardware failure. MCU not inserted properly in the
socket, extract MCU, and verify pins not to be shorted.
Damaged ADC converter, replace PCB. If a PCB
adapter is used, verify that PCB adapter is properly
secured in the socket and that there are no bent pins. If
PCB rev 2.x is in place make sure that C40 is removed.
“NO
CONFIGURATION
EEPROM”
Lack of communication between microcontroller and
EEPROM
Hardware failure. MCU not inserted properly in the
socket, extract MCU, and verify pins not to be shorted.
Damaged or wrong inserted or missing EEPROM,
replace software kit. Damaged controller PCB. If a PCB
Adapter is used, verify that PCB adapter is properly
secured in the socket and that there are no bent pins