9. Statim 2000 Trouble-shooting
Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx
37
Cycle Fault #98
Software Revisions
R4xx and up
(includes R5xx and
R6xx)
Microcontroller fails to communicate with ADC
(Analog to Digital Converter)
Hardware failure. MCU not inserted properly in
the socket, extract MCU, and verify pins not to
be shorted. Damaged ADC converter, replace
PCB. If a PCB adapter is used, verify that PCB
adapter is properly secured in the socket and
that there are no bent pins. If PCB rev 2.x is in
place make sure that C40 is removed.
“NO
CONFIGURATION
EEPROM”
Lack of communication between microcontroller and
EEPROM
Hardware failure. MCU not inserted properly in
the socket, extract MCU, and verify pins not to
be shorted. Damaged or wrong inserted or
missing EEPROM, replace software kit.
Damaged controller PCB. If a PCB Adapter is
used, verify that PCB adapter is properly
secured in the socket and that there are no bent
pins