Technical Data
Operating Manual ProNumeric XCI 600 Version 07/16
93
Climatic conditions
Ambient operating temperature
0 ... +50°C (free air circulation)
Storage temperature
-25 ... +70°C
Relative humidity
20 ... 80%, non condensing
Air pressure in operation
860 ... 1060 hPa
Mechanical strength
Vibrations
Acc. to DIN EN 60068-2-6
9 ... 150 Hz constant acceleration 1 g
10-500Hz, 1g
Shock
Acc. to DIN EN 60068-2-27, sinusoidal half-wave 15g / 11ms
Free fall
Acc. to DIN EN 60068-2-32, drop height 1m (in original packaging)
Electrical safety
Protection class
Note for the operator:
When connected to the mains, the device must be connected via a safety
transformer (reliable isolation)
Protection type
IP 20 to EN 60529
Air gap/creepage paths
according to EN60950-1
Test voltage
--
Electromagnetic compatibility
Electrostatic discharge
EN 61000-4-2: 8 kV air discharge, 4 kV contact discharge
Electromagnetic fields
EN 61000-4-3: Field intensity 10 V/m, 80...1000 MHz
Rapid transients
(bursts)
1 kV on I/O signal lines
Energetic transients
(surges)
EN 61000-4-5:
0.5 kV DM (differential mode)
1 kV CM (common mode) on shielded signal lines
Interference emissions
EN 55011: Limit category A, Group 1