STB8827
Series Operation Manual Chapter 3 Introduction to [MEAS]
Note:
The correction function can influence the level monitor function, so when
the correction data changes the level monitor value will change. When the
correction is switched between OPEN or SHOR or LOAD, the level monitor value
will be influenced as well
———————————————————————————————————————
Operation steps for setting the level monitor function
Execute the following operation steps to set the level monitor function as ON or
OFF.
1) Move the cursor to
Vm
28
zone, the following soft keys will be displayed.
ON
OFF
2) Press
ON
to set the voltage monitor function as ON while press
OFF
to set the
voltage monitor function as OFF.
3) Move the cursor to
Im
zone, the following soft keys will be displayed.
ON
OFF
4) Press
ON
to set the current level monitor function as ON while press
OFF
to
set the current level monitor function as OFF.
3.5.6 Delay time
STB8827
trigger delay means the delay time from triggering to test-start.
Delay function can set the trigger delay time. When the list sweep test function is
used, all set delay time will be delayed at each sweep test point.
The range of the trigger
delay time can be set from 0s to 60s with 1ms as the resolution. The trigger delay
function is great useful when the instrument is applied in an auto test system.
When the instrument is triggered by HANDLER interface, the trigger delay time
can ensure DUT and test terminal has a reliable contact.
Operation steps for setting the delay function
Execute the following steps to set the measurement delay time.
1) Move the cursor to the
DELAY
zone.
2) Use numeric keys to input delay time. After pressing a numeric key, the
following unit keys will be displayed. These soft keys can replace
[ENTER]
to
input delay time.
msec
sec
3.5.7 Output impedance
STB8827
provides four output impedances for your choice: 100
Ω
, 50
Ω
and 30
Ω
. When testing inductance, it is necessary to input the same output impedance
so as to make data comparison with other instruments.