Series Operation Manual Chapter 3 Introduction to [MEAS]
3.1.2
15
Test range
Measurement range should be selected in accordance with the impedance value
of the tested LCR component.
STB8827
has 9 AC measurement ranges: 10
Ω
, 30
Ω
, 100
Ω
, 300
Ω
, 1k
Ω
, 3k
Ω
, 10k
Ω
,
30k
Ω
, 100k
Ω
.
STB8827
has 13 DCR measurement ranges: 100m
Ω
, 300m
Ω
, 1
Ω
, 10
Ω
, 100
Ω
,
300
Ω
, 1k
Ω
, 3k
Ω
, 10k
Ω
, 30k
Ω
, 100k
Ω
, 300k
Ω
, 1M
Ω
.
Operation steps for setting test range:
1) Move the cursor to the range zone, the following soft keys will be displayed:
AUTO The soft key is used to set the range mode to AUTO.
HOLD The soft key is used to switch the AUTO mode to the HOLD mode.
When the range mode is set to HOLD, the range will be locked in the
current measurement range. The current measurement range will be
displayed in the range zone.
↑
(+) The soft key is used to increase the range under HOLD mode.
↓
(-) The soft key is used to decrease the range under HOLD mode.
2) Use soft keys to set measurement range.
3.1.3 Test frequency
The measurement range of
STB8827
ranges from 20Hz to 1MHz with an
increase or decrease of 0.01Hz. When the test function is set as DCR, the
FREQ
zone will display “---”.
Frequency of
STB8827
A: 20Hz~300kHz
Frequency of
STB8827
B: 20Hz~500kHz
Frequency of
STB8827
C: 20Hz~1MHz
Frequency range and frequency range points
Frequency range (F)
Test frequency point
Resolution
20Hz
≤
F
≤
99.99Hz
20.00Hz, 20.01Hz ……99.99Hz 0.01Hz
100Hz
≤
F
≤
999.9Hz
100.0Hz, 100.1Hz ……999.9Hz
0.1Hz
1kHz
≤
F
≤
9.999kHz
1.000kHz, 1.001kHz ……9.999kHz 1Hz
10kHz
≤
F
≤
99.99kHz 10.00kHz, 10.01kHz……99.99kHz
10Hz
100kHz
≤
F
≤
1MHz
100.0kHz, 100.1 kHz……1MHz
100Hz
Operation steps for setting test frequency:
STB8827
provides two methods to set measurement frequency. The first one is
to use soft keys and the other one is to input data by using numeric keys.
1) Move the cursor to the
FREQ
zone, the following soft keys will be displayed.
↑
(++)
This is a coarse adjustment soft key used to increase the frequency. Press