Series Operation Manual Chapter 3 Introduction to [MEAS]
36
value should conform with the preset load correction test function.
Load correction function adopts the transport coefficient between the real
test value of preset frequency and the standard reference value to eliminate the
test error. Load correction function is only available for calculating transport
coefficient.
3.6.5 Cable length selection
The available cable length is 0m, 1m, 2m and 4m.
3.7 <LIMIT TABLE>
Press
[SETUP]
and then
LIMIT TABLE
to enter into the
<LIMIT TABLE
SETUP>
page as the following figure shown.
STB8827
307
Compare function can be set on this page.
STB8827
can set 9 bin limits of
primary parameters and one of secondary parameters. The tested result can be
divided into up to 10 bins (BIN 1 to BIN 9 and BIN OUT). If the primary
parameter of DUT is within the limit range from BIN1 to BIN9,
but the secondary parameter is out of the limit range, in this case the DUT will
be sorted into aux bin. When
STB8827
installs the HANDLER interface and it is
used in automatic sorting system, the compare function will be especially
useful. The following limit parameter of compare function only be set on
<
LIMIT TABLE SETUP>
page.
Test parameter (
PARAM
)
Limit mode of compare function (
MODE
)
Nominal value (
NOM
)
Auxiliary bin ON/OFF (
AUX
)
Compare function ON/OFF (
COMP
)