STB8827
Series Operation Manual Chapter 7 Command Reference
ABSolute
FUNCtion:DEV<n>:MODE PERCent
OFF
Where,
ABSolute Absolute value deviation display
PERCent Percent deviation display
OFF Real value display
Where, <n> is
Character 1 (49) is equal to the nominal value of primary parameter.
Or character 2 (50) is equal to the nominal value of the secondary parameter.
90
For example: WrtCmd (“FUNC:DEV1:MODE ABS”)
Query syntax: FUNCtion:DEV<n>:MODE?
Return format:
ABX
PERC <NL^END>
OFF
The FUNCtion:DEV<n>:REFerence<value> command is used to set the nominal
value of the deviation. The FUNCtion:DEV<n>:REFernece<value>? query
returns the current nominal value of the deviation.
Command syntax: FUNCtion:DEV<n>:REFerence<value>
Where,
<value> is NR1, NR2 or NR3 data format.
<n> is
Character 1 (49) is equal to the nominal value of primary parameter.
Or Character 2 (50) is equal to the nominal value of the secondary parameter.
For example: WrtCmd (“FUNC:DEV1:REF 10”)
Query syntax: FUNCtion:DEV<n>:REFerence?
Return format
:
<NR3><NL^END>
The FUNCtion:DEV<n>:REFerence:FILL command is used to set the nominal
value of the deviation. This command directs the instrument to make a test and
then copies the results of the primary and the secondary parameters as the
nominal values of the deviation.
Command syntax: FUNCtion:DEV<n>:RERerence:FILL
Where,
Character 1 (49) or character 2 (50) is equal to the nominal values of the