STB8827
Series Operation Manual Chapter 3 Introduction to [MEAS]
frequency is displayed on the frequency setting zone.
3)
35
Use numeric keys to input the correction frequency. After pressing any
numeric key, the available unit keys (
Hz
,
kHz
and
MHz
) will be displayed on
the soft key zone and these soft keys can replace the
[ENTER]
key to input
correction frequency. When the
[ENTER]
key is used to input correction
frequency, the default unit is Hz.
4) Connect the test fixture to the test terminal.
5) Make the test fixture be open.
6) Press
MEAS OPEN
to perform open correction at the current set frequency.
The test result (G, B) of the open correction test will be displayed in the help
line (the bottom line).
7) Move the cursor to
OPEN
.
8) Press
ON
to perform the open correction calculation at preset frequency in
latter measurements.
9) Move the cursor to
FREQ
to set the required correction frequency.
10) Make the test fixture be short.
11) Press
MEAS SHORT
to perform short correction at preset frequency. The test
result (R, X) of the short correction will be displayed in the help line (the
bottom line).
12) Move the cursor to
SHORT
.
13) Press
ON
to perform the short correction calculation at preset frequency in
latter measurements.
14) Prepare a standard test component.
15) Move the cursor to
FUNC
.
16) Set the function parameters required to be set.
17) Move the cursor to
REF A
.
18) Use numeric keys and unit keys to input the primary reference values of the
standard component.
19) Move the cursor to
REF B
.
20) Use numeric keys and unit keys to input the secondary reference value of the
standard component.
21) Move the cursor to the corresponding
FREQ
.
22) Connect the standard component to the test fixture.
23) Press
MEAS LOAD
, the instrument will execute a load correction. The real
test results of the standard component will be displayed in
LOAD A
and
LOAD B
.
24) Move the cursor to
LOAD
.
25) Press
ON
to perform load correction calculation at preset frequencies in latter
measurements.
3.6.4 Load correction test function
When performing load correction, the reference value of the standard
component is required to be input in advance. The test parameters of reference