5) Multi-marker and Marker Table
The instrument can display 6 markers simultaneously. After pressing “Select Mkr” soft-key, User can switch
between these markers and set the “On/Off” state of it.
After selecting the marker and set the state of it to “On”, user can change the marker’s position by the same way of
changing “Normal Mkr”.
Fig.3.57 Marker table on
NOTE:
When switch between “Mkr Normal”, “Delta Mark”, “Fixed Mark” and multi-markers, the pre-marker
will automatically set to “Off”.
6) All off
Pressing "All Off" soft-key, all markers will be set to “Off”.
7) Count
Frequency count can give the accurate frequency of the strongest signal around the marker, which is working as a
cymometer. The bandwidth of the counter is +/- 100kHz and the accuracy of it is Hz.
NOTE:
If the “Count” soft-key is set to “On”, the displayed frequency will change from the frequency of marker
to the frequency of the strongest signal around the marker.
8) Mark Noise
Set “Mark Noise” to “On”, and then user can get the average power of the noise. It will take the place of the
amplitude to display on the up-right of the screen.