Fig.3.37 Sample detector mode
2) Positive-Peak
Detector sample the positive value of VBW filter output of each sweep point, and display it. If the sweep time is long
enough, this detector can put out the maximum envelop of the modulated signal.
Fig.3.38 Positive-peak detector mode
3) Negative-Peak
This detector is opposite to “Posi-Peak” detector. The output of it is the minimum envelop of the signal.